Exchanging the Outlet Valve

When
If internally leaking
Tools required
p/n
8710-0510
8710-1924
Parts required
p/n
G1312-60067
Preparations
• Switch off pump at the main power switch
• Remove the front cover
1
Using a 1/4 inch wrench disconnect the valve capillary
from the outlet valve. Using the 14 mm wrench loosen the
valve and remove it from the pump body.
Valve capillary
Outlet valve
1260 Infinity QuatPump VL User Manual
Description
Wrench open 1/4 — 5/16 inch
Wrench open 14 mm
Description
Outlet valve 1220/1260
2
Do not disassemble the outlet valve, as this can damage
the valve.
Maintenance

Exchanging the Outlet Valve