Exchanging The Outlet Ball Valve - Agilent Technologies 1100 Series Reference Manual

Quaternary pump
Hide thumbs Also See for 1100 Series:
Table of Contents

Advertisement

When required
Tools required
Parts required
N O T E
1 Using a 1/4 inch wrench disconnect the valve capillary from the outlet ball
2 Using the 14 mm wrench loosen the valve and remove it from the pump body.
3 Check that the new valve is assembled correctly and that the gold seal is
Figure 15
Valve body
Gold seal
Plastic cap
4 Reinstall the outlet ball valve and tighten the valve.
5 Reconnect the valve capillary.
Repairing the Pump

Exchanging the Outlet Ball Valve

Exchanging the Outlet Ball Valve

If internally leaking
Wrench 1/4 inch
Wrench 14 mm
Outlet ball valve G1311-60012
Before exchanging the outlet ball valve you can try to clean it in a sonic bath.
Place the valve in upright position (onto the plastic cap) in a small beaker with
alcohol. Place in a sonic bath for 5 – 10 minutes.
valve.
present (If the gold seal is deformed, it should be replaced).
Outlet Ball Valve Parts
106