9
Maintenance and Repair

Exchange Flush Pump Outlet Valve

Exchange Flush Pump Outlet Valve

When
If the Flush pump delivery test fails or if the Outlet valve show signs of malfunction.
Tools required
• 1/4 inch X 5/16 inch wrench (p/n 8710-0510)
• Wrench open 14 mm (p/n 8710-1924)
Parts required
#
5067-4716
1
Open the front door of the module.
100
Description
Outlet valve
2
With a ¼ inch wrench disconnect the capillary at the top
of the Outlet valve.