Calibration Curves - AB Sciex 4500 Series System User's Manual

Table of Contents

Advertisement

Figure 8-10 Peak Review Right-Click Menu
Menu
Options
Sample Annotation
Save Active to Text File
Show First Page
Show Last Page
Slide Show Peak Review
Update Method
Revert to Method
Delete Pane

Calibration Curves

Use calibration curves to find the calculated concentration of samples, including QC (quality
control) samples. QC samples are added to a batch to estimate the data quality and accuracy of
standards in the batch. QC samples have known analyte concentrations but are treated as
unknowns so that the measured concentrations can be compared to the actual value.
The calibration curve is generated by plotting the concentration of the standard versus its area or
height. If an internal standard is used, the
versus the ratio of the standard peak height or area to the internal standard peak height or area is
plotted. The area or height ratio of a sample is then applied to this curve to find the concentration
System User Guide
Document Number: D5030863 A
Function
Opens the Peak Review Options dialog.
Opens the Sample Annotation dialog.
Saves the selected peak as a text file.
Goes to the first sample.
Goes to the last sample.
Opens the slide show.
Updates the algorithm for all peaks.
Click to have a redefined peak reselected based on the current
quantitation method.
Deletes the active pane.
ratio of the
Analyzing and Processing Quantitative Data
standard concentration/internal standard
4500 Series of Instruments
97 of 142