Figure 7-2
3. In the Mass Tolerance field, type the value to dictate the mass range used to find a
peak. The software finds the peak using a value twice the typed range (± the mass
value).
4. In the Minimum Intensity field, type the intensity below which peaks are ignored by
the algorithm.
5. In the Minimum Mass field, type the mass that determines the beginning of the scan
range.
6. In the Maximum Mass field, type the mass that determines the end of the scan
range.
7. To set the start and end times, select the Use Limited Range check box and do the
following:
In the Start Time field, type the time that determines the start of the
experiment.
In the End Time field, type the time that determines the end of the experiment.
8. Click OK.
The BPC is generated in a new pane.

Generate XWCs

The user can extract up to three ranges from a DAD spectrum to generate the XWC. For more
information about using the available icons, refer to
1. Open a data file that contains a DAD spectrum.
2. Anywhere in the pane, right-click and then click Extract Wavelengths.
System User Guide
Document Number: D5030863 A
Base Peak Chromatogram Options dialog
Analyzing and Processing Data
Table 7-8 Graph Options on page
4500 Series of Instruments
81.