Analyzing and Processing Quantitative Data
Figure 8-6
Item Description
1
Arrows: click to move through the peaks.
2
Show or Hide Parameters: click to show the integration parameters.
3
Integration parameters: click to change the parameters.
10. Click Show or Hide Parameters twice.
11. Click Specify Parameters -MQ III.
12. Change the Noise Percent value.
13. Click Apply.
The peak is integrated closer to the baseline.
14. If the change does not improve the peak integration, then adjust the Noise Percent
parameter until the optimal value is found.
15. To update the algorithm for all peaks, right-click in the pane and then click Update
Method.
Note: The Update Method function only updates the algorithm values for
that specific analyte (or internal standard) and not all analytes.
4500 Series of Instruments
94 of 142
2
1
3
Peak Review Pane
System User Guide
Document Number: D5030863 A