Configure the Integrated Syringe Pump

1. In the Navigation bar, under Acquire, double-click Build Acquisition Methods.
2. In the Acquisition method pane, click the Syringe Pump icon.
The Syringe Pump method properties tab opens in the Acquisition Method Editor
pane.
3. In the Syringe Diameter (mm) field, type the syringe diameter.
4. In the Flow Rate field, type the flow rate.
5. In the Unit list, select the units of flow.
6. Save the file.

Configure the Diverter Valve

The diverter valve is a two position, 6 port valve. If the valve is in Position A
valve—injector mode position
switched to Position B
injected. For more information about creating and editing hardware profiles, refer to
Hardware Profile on page
1. In the Navigation bar, under Configure, double-click Hardware Configuration.
2. Create or edit the hardware profile containing the instrument.
3. In the Configuration tab
injector/diverter valve.
System User Guide
Document Number: D5030863 A
Syringe pump LED
Syringe pump on/off button
Syringe pump status LED
A), the sample flows through the external loop. When the valve is
(Figure 2-10 Diverter valve—injector mode position
31.
(Figure 2-8 Configuration
1
2
(Figure 2-9 Diverter
B), the sample is
tab), select Use integrated