The following figure shows the location of the parameters on the ion optics path.
1
Figure 5-1 Ion optics path and parameters
Item Parameter
1
IS (Ion Transfer
Voltage)
1
NC (Needle
Current)
1
ihe (Interface
Heater)
1
IHT (Interface
Heater
Temperature)
1
sdp
System User Guide
Document Number: D5030863 A
2
Parameter
Use
Type
Source and
For the PhotoSpray
gas
parameter controls the voltage that
transfers the ions from the primary
ionization region towards the
curtain plate orifice.
Source and
The NC parameter controls the
gas
current applied to the corona
discharge needle in the APCI
probe, used in the Turbo V™
source. The discharge ionizes
solvent molecules, which in turn
ionize the sample molecules.
Source and
The ihe parameter switches the
gas
interface heater on and off. Heating
the interface helps maximize the
ion signal and prevents
contamination of the ion optics.
This should always stay on. The
button controlling the interface
heater reads ON when the
interface heater is on.
Source and
The IHT parameter controls the
gas
temperature of the NanoSpray
interface heater and is only
available if the NanoSpray ion
source and interface are installed.
Source and
The sdp parameter controls the
gas
selection of the DuoSpray™ ion
source probes: TurboIonSpray