AB Sciex 4500 Series System User's Manual page 126

Table of Contents

Advertisement

4500 Series Instrument Requirements and Parameters
Table B-3 4500 Series Instrument Parameters
Access ID
Positive ion mode
Q1
(6)
CUR
20
10 to 55
(1)(2)(3)(4)
CUR
10
(5)
10 to 55
(a)
CAD
0
Fixed
(b)
CAD
0
Fixed
(1)(2)
IS
5500
0 to 5500
(5)
IS
1500
Ion Transfer
0 to 2500
Voltage
(6)
IS
1000
0 to 4000
(3)(4)
NC
3
0 to 5
(2)(3)(4)(5)
TEM
0
0 to 750
(a)
DP
200
0 to 300
(b)
DP
130
0 to 300
EP
10
2 to 15
(a)
CEM
1800
0 to 3300
(b)
CEM
2000
0 to 3300
(1)IonSpray™ ion source (2)TurboIonSpray
(4)DuoSpray™ ion source, 1=TIS, and 2=HN (5)PhotoSpray
ion source (a) AB SCIEX QTRAP
4500 Series of Instruments
126 of 142
Q3
MS/MS
20
20
10 to 55
10 to 55
10
10
10 to 55
10 to 55
6
Medium
Fixed
Low,
Medium,
High
5
9
Fixed
0 to 12
5500
5500
0 to 5500
0 to 5500
1500
1500
0 to 2500
0 to 2500
1000
1000
0 to 4000
0 to 4000
3
3
0 to 5
0 to 5
0
0
0 to 750
0 to 750
100
100
0 to 300
0 to 300
130
120
0 to 300
0 to 300
10
10
2 to 15
2 to 15
1800
1800
0 to 3300
0 to 3300
2000
2000
0 to 3300
0 to 3300
®
ion source (3)Heated Nebulizer
®
4500 system (b) AB SCIEX Triple Quad™ 4500 system
Negative ion mode
Q1
Q3
20
20
10 to 55
10 to 55
10
10
10 to 55
10 to 55
0
6
Fixed
Fixed
0
5
Fixed
Fixed
–4500
–4500
–4500 to 0
–4500 to 0
–1500
–1500
–2500 to 0
–2500 to 0
–1000
–1000
–4000 to 0
–4000 to 0
–3
–3
–5 to 0
–5 to 0
0
0
0 to 750
0 to 750
–100
–100
–300 to 0
–300 to 0
–60
–60
–300 to 0
–300 to 0
–10
–10
–15 to –2
–15 to –2
1800
1800
0 to 3300
0 to 3300
2000
2000
0 to 3300
0 to 3300
®
ion source (6) NanoSpray
System User Guide
Document Number: D5030863 A
MS/MS
20
10 to 55
10
10 to 55
Medium
Low,
Medium,
High
9
0 to 12
–4500
–4500 to 0
–1500
–2500 to 0
–1000
–4000 to 0
–3
–5 to 0
0
0 to 750
–100
–300 to 0
–150
–300 to 0
–10
–15 to –2
1800
0 to 3300
2000
0 to 3300
®

Hide quick links: