Edit The Hardware Profile For The Integrated Syringe Pump; Adjust The Integrated Syringe Pump Position - AB Sciex 4500 Series System User's Manual

Table of Contents

Advertisement

System Information
Edit the Hardware Profile for the Integrated
Syringe Pump
Make sure the syringe pump is seated properly to avoid damaging the syringe. For more
information about creating and editing hardware profiles, refer to
page
31.
1. In the Navigation bar, under Configure, double-click Hardware Configuration.
2. Create or edit the hardware profile containing the instrument.
3. In the Configuration tab
syringe pump.
4. Click OK.
Figure 2-3
5. Activate the hardware profile.

Adjust the Integrated Syringe Pump Position

1. Press the button on the right side of the syringe pump to lower the base and then
insert the syringe as shown in
Make sure that the end of the syringe is flush against the base and that the shaft of
the syringe rests in the cutout.
4500 Series of Instruments
22 of 142
(Figure 2-3 Configuration
Configuration tab
Figure 2-4 Lowering the
Create a Hardware Profile on
tab), select Use integrated
syringe.
System User Guide
Document Number: D5030863 A