Analyzing and Processing Quantitative Data
Figure 8-8
Item Description
1
2. Zoom in on the lower 10% of the peak.
Figure 8-9
Item
1
3. Position the cross-hair where the start of the peak is to be defined and then drag the
cross-hair to where the end of the peak is to be defined.
The software shades the area bounded by the base and sides of the peak. Peak
parameters are gray as they are no longer applicable because the peak was drawn
manually.
4. Do one of the following:
To make this change permanent, click Accept.
To discard the changes, clear the Manual Integration check box.
Tip!
then click Revert to Method.

Peak Review Right-Click Menu

Right-click in the Peak Review window or pane to access the following options.
4500 Series of Instruments
96 of 142
Peak Review pane: Manual integration
Manual Integration Mode
1
Peak Review pane: Zooming in on a peak
Description
Lower 10% of the peak
If a peak was correct as originally selected, right-click the peak and
System User Guide
Document Number: D5030863 A