Basic Acquisition Methods

Copy an Experiment into a Period

Prerequisite: Multi-period method
In the Acquisition method pane, press CTRL, and then drag the experiment to the
period.
The experiment is copied below the last experiment in the period.

Copy an Experiment within a Period

Use this procedure to add the same or similar experiments to a period if most or all of the
parameters are the same.
Right-click the experiment and then click Copy this experiment.

Scan Techniques

MS: In MS scans, also referred to as single MS scans, ions are separated according to their
mass-to-charge ratio. A single MS scan may be used to determine the molecular weight of a
compound. Single MS scans can also be referred to as survey scans. MS scans do not provide
any information as to the chemical make-up of the ions other than the mass-to-charge ratio.
Perform MS/MS or MS/MS/MS scan types to obtain more information about the ions.
MS/MS: MS/MS scans are used to help identify or confirm a molecular species. In MS/MS, a
precursor ion can be fragmented in one of two locations.
For triple quadrupole instruments, fragmentation occurs in the collision cell.
For LIT instruments, fragmentation can occur in the collision cell or the linear ion
trap.
If enough energy is used, the precursor ion fragments to produce characteristic product ions.
MS/MS/MS: The LIT instrument MS/MS/MS scans go one step further than MS/MS scans. A
fragment that is produced in the collision cell is fragmented further in the trap to give more
structural information about the molecular ion.

Quadrupole-Mode Scan Types

Triple quadrupole instruments have high-sensitivity MRM (Multiple Reaction Monitoring)
capabilities required for quantitation experiments. In addition, they have highly specific scan
types such as precursor ion and neutral loss scans that allow a more advanced search to be
performed on the components of the samples.
Q1 MS (Q1): A full scan using the first quadrupole (Q1). The ion intensity is returned for every
mass in the scan range.
Q1 Multiple Ions (Q1 MI): A zero width scan type using the first quadrupole (Q1). The ion
intensity is returned for the specified masses only.
Q3 MS (Q3): A full scan using the third quadrupole (Q3). The ion intensity is returned for every
mass in the scan range.
Q3 Multiple Ions (Q3 MI): A zero width scan type using the third quadrupole (Q3). The ion
intensity is returned for the specified masses only.
4500 Series of Instruments
44 of 142
System User Guide
Document Number: D5030863 A