Figure 8-7

Manually Integrate Peaks

Manually integrating peaks should be done last. Manually integrate peaks only if all the peaks
have not been found after adjusting and updating the algorithm parameters. This is done to limit
person-to-person variability.
Note: Peaks that are manually integrated, or where the algorithm was changed for
only that peak, are identified as such in the Record Modified column of the Results
Table, as are the peaks that have algorithm parameter changes for a sample but not
updated to the entire analyte group.
1. In the Peak Review pane, click Manual Integration Mode.
System User Guide
Document Number: D5030863 A
Update Method
1
Analyzing and Processing Quantitative Data
4500 Series of Instruments
95 of 142