14. In the Sample tab, in the Acquisition section, select a method from the list.
Depending on how the system is set up, specific information for the autosampler
must be entered. Even if the injection volume is set in the method, the injection
volume can be changed for one or more samples by changing the value in the
injection volume column.
Tip!
the Use Multiple Methods check box. The Acquisition Method column is
shown in the Sample table. Select the acquisition method for each sample
in this column.
15. To change the injection volumes from the volumes listed in the method, in the Inj.
Volume (µL) column, type the injection volume for each sample.
16. To set sample locations, do one of the following:
Set Sample Locations in the Batch Editor on page 59
Select Vial Positions using the Locations Tab on page 59
17. Click the Submit tab.
18. If the Submit Status section contains a message about the status of the batch, do
one of the following:
If the message indicates that the batch is ready for submission, proceed to
step 19.
If the message indicates that the batch is not ready for submission, make the
changes as indicated by the message.
19. Click Submit.

Change Sample Order

The order of the samples can be edited before they are submitted to the Queue.
In the Submit tab, double-click any of the numbers on the far left of the table (a very
faint square box is visible), and then drag them to the new location

Acquire Data

The system should not be in Tune mode when sample acquisition is started. Also, if the system
has been previously run that day and has not yet been set to Standby, sample acquisition will
start automatically.
1. In the Navigation Bar, click Acquire.
2. Click View > Sample Queue.