Advertisement

3. Make sure that the mains supply cable is plugged in to the instrument.
4. Make sure the instrument and roughing pump mains supply cable are plugged into
the 200 to 240 V electrical mains supply.
5. Make sure that the Ethernet cable is connected to both the instrument and the
computer.
6. Turn on the roughing pump. The mains switch is located beside the mains supply
input connection on the roughing pump.
7. Wait 5 minutes, then turn on the instrument switch.
Note: If the switch is already on, turn it off and then on.
8. Turn on the computer, if it was turned off.
9. Start the software.

Reset the System

Press and hold the Reset button for 5 seconds.
An audible click is heard when the relay activates. After approximately three
minutes, the instrument should reach operating pressure.

Shut Down the System

1. Complete or stop any ongoing scans. For more information, refer to
Acquisition on page
2. Shut off the sample flow to the instrument.
Caution: Potential Instrument Damage: Shut off the sample flow before the
instrument is shut down.
3. In the Analyst software, deactivate the hardware profile, if active.
4. Close the software.
5. Press and hold the Vent button for three seconds.
The turbo pump will spin down gradually.
6. Wait 15 minutes.
7. Turn off the roughing pump.
The mains switch is located beside the mains supply input connection on the
roughing pump.
8. Wait 15 minutes, then turn off the instrument switch.
9. Unplug the mains supply cable from the bulkhead on the left side of the instrument.
10. Unplug the roughing pump mains supply cable from the electrical mains supply.