Manually Integrate Peaks - AB Sciex QTRAP 4500 LC/MS/MS system System User's Manual

Qtrap 4500 series
Table of Contents

Advertisement

Figure 11-7 Update Method

Manually Integrate Peaks

Manually integrating peaks should be done last, to limit person-to-person variability. Manually
integrate peaks only if all the peaks have not been found after the algorithm parameters have
been adjusted and updated.
Note: Peaks that are manually integrated, or where the algorithm was changed for
only that peak, are identified as such in the Record Modified column of the Results
Table, as are peaks that have algorithm parameter changes for a sample that are not
applied to the entire analyte group.
1. In the Peak Review pane, click Manual Integration Mode.
Figure 11-8 Peak Review Pane: Manual Integration
System User Guide
RUO-IDV-05-0264-A
Operating Instructions — Analyze and Process Quantitative Data
1
4500 Series of Instruments
105 of 138