Leakage test

The following tests are initiated during the leakage
test:
– Compliance Test
– System Leak Test
– Ventilator Leak Test
– Safety Relief Valves Test
Standby
1 Press the Leak /ComplTest softkey (A).
2 Follow the instructions on the screen.
Standby
After completion of the tests, the results (B) are
displayed on the screen.
3 To return to the start screen, press the rotary
knob.
Instructions for use Fabius plus SW 3.n
A
B
Results of the compliance test
This test determines the system compliance
including breathing system, breathing hoses, filter,
and Y-piece.
The system compliance is required to ensure in
Volume Control mode that the applied tidal
volume corresponds to the set tidal volume.
System compliance
[mL/cmH
O]
2
≤6.5
The value of the compliance is displayed on the
Standby screen.
Results of the ventilator leakage test
The test of the ventilator leakage can have the
following results:
Ventilator leakage
[mL/min]
150
151 to 250
>250
Configuration
Displayed result
[mL/cmH
O]
2
Measured value and
PASSED
Displayed result
[mL/min]
Measured value and
PASSED
Measured value and
FAILED
>250 and FAILED
135