Parameter field for O
Volume Control
The following information is displayed in the
parameter field for O
monitoring:
2
A Numerical value for the inspiratory O
concentration in percent (%) in the range from
10 % to 100 %
B Upper alarm limit for the O
(%)
C Lower alarm limit for the O
(%)
Instructions for use Fabius plus SW 3.n
monitoring
2
B
A
C
2
concentration in
2
concentration in
2
Setting the O
alarm limits
2
The standard alarm limits configured for the
ventilation mode can be used unchanged, see
chapter ''Restoring the factory settings'' on page
142.
or
The alarm limits can be set individually for the
current case:
1 Press the
(A) key.
Volume Control
The dialog window with the alarm limits (B) opens.
2 Adjust the upper and lower alarm limit values of
the O
concentration (C), see setting ranges of
2
the alarm parameters in chapter ''Changing the
alarm limits'' on page 141.
3 Confirm new values.
Monitoring
A
C
B
123