Recommended safety clearance for portable and mobile high-frequency
communication equipment
The safety clearances listed in the following
comply with IEC 60601-1-2.
Max. P
150 kHz to 2.5 GHz
EIRP
(watts)
0.03
0.32 m (1.1 ft)
0.10
0.58 m (1.9 ft)
0.17
0.76 m (2.5 ft)
0.20
0.82 m (2.7 ft)
0.25
0.92 m (3.0 ft)
0.41
1.2 m (3.9 ft)
0.82
1.7 m (5.6 ft)
1.00
1.8 m (5.9 ft)
1.64
2.4 m (7.9 ft)
3.3
3.3 m (11 ft)
Reduced safety clearance for portable
and mobile high-frequency
communication equipment
The safety clearances listed in the following are
the result of tests that Dräger has performed to
determine the minimum necessary safety
clearances. These reduced safety clearances
apply only to mobile high-frequency
communication equipment that uses the standards
specified.
Mobile high-frequency communication equipment with ...
GSM 850, GSM 900, RFID 868 MHz (limited to 2 W ERP)
GSM 1800, GSM 1900 (limited to 1 W ERP)
UMTS, DECT (limited to 0.25 W ERP)
Bluetooth, WLAN 2450, RFID 2450 (limited to 0.1 W ERP)
Instructions for use Fabius plus SW 3.n
All other frequen-
Examples
cies
0.96 m (3.2 ft)
WLAN 5250 / 5775 (Europe)
1.8 m (5.9 ft)
WLAN 2440 (Europe)
2.3 m (7.6 ft)
Bluetooth, RFID 2.5 GHz
2.5 m (8.2 ft)
WLAN 5250 (not in Europe)
2.8 m (9.2 ft)
UMTS mobiles
3.5 m (12 ft)
Cordless DECT devices
5.0 m (16 ft)
RFID 13.56 MHz
5.5 m (18 ft)
WLAN 5600 (not in Europe)
7.1 m (23 ft)
GSM 1800 / GSM 1900
10 m (33 ft)
GSM 900 mobile phones, RFID 868 MHz
Technical data
Safety clearance
0.30 m (12 in)
0.30 m (12 in)
0.15 m (6 in)