Leakage Test - Dräger Fabius Tiro Instructions For Use Manual

Anesthesia workstation
Hide thumbs Also See for Fabius Tiro:
Table of Contents

Advertisement

Configuration
Troubleshooting with failed O
 Replace the O
sensor capsule in the O
2
sensor housing, see chapter ''Inserting a new
O
sensor capsule'' on page 75.
2
If the calibration continues to fail, contact
DrägerService or the authorized local service
partner.

Leakage test

The following tests are initiated during the leakage
test:
– Compliance Test
– System Leak Test
– Ventilator Leak Test
– Safety Relief Valves Test
Standby
1 Press the Leak /ComplTest softkey (A).
128
calibration
2
2
A
2 Follow the instructions on the screen.
Standby
After completion of the tests, the results (B) are
displayed on the screen.
3 To return to the start screen, press the rotary
knob.
Results of the compliance test
This test determines the system compliance
including breathing system, breathing hoses, filter,
and Y-piece.
The system compliance is required to ensure in
Volume Control mode that the applied tidal
volume corresponds to the set tidal volume.
System compliance
[mL/cmH
O]
2
≤6.5
The value of the compliance is displayed on the
Standby screen.
Instructions for use Fabius Tiro SW 3.n
B
Displayed result
[mL/cmH
O]
2
Measured value and
PASSED