Suction systems used
When using an ejector suction with drive gas:
 Connect the Air connection hose of the suction
system via a 3-way adapter to the gas supply
block or the central gas supply.
For vacuum suction:
 Connect the vacuum hose of the suction
system directly to the central gas supply.
Prepare the suction system in accordance with the
associated instructions for use.
WARNING
Risk of patient injury
Only use the suction system if the ventilation
mode ManSpont is active or the patient is
disconnected from the Y-piece.
Instructions for use Fabius plus SW 3.n
Assembly and preparation
71