Assembling the breathing system

WARNING
Risk of insufficient anesthetic gas
concentrations
If the component connections of the
breathing system are not leak-tight enough,
ambient air may get into the breathing gas.
Make sure that all components of the
breathing system are connected tightly.

Preparing the ventilator

D
C
B
A
Instructions for use Fabius plus SW 3.n
F
G
E
Only disinfected and sterilized components must
be used.
1 Open the ventilator door (A) with the attached
ventilator unit.
2 Release the 3 clamps (B).
3 Remove the cover (D).
4 Insert the ventilator membrane (C). After
assembly, the Dräger inscription must be
visible.
5 Fit the cover (D) and close the 3 clamps (B).
6 Connect the pressure sensor line (E) of the
ventilator chamber with the appropriate
connector.
7 Close the ventilator door with the attached
ventilator unit.
Safety functions of the ventilator
– Overpressure safety valve (F)
– Underpressure safety valve (G)
– Pressure sensor in the ventilator chamber
Assembly and preparation