Ventilator Failure; Alarm Ventilator Fail - Dräger Fabius plus Instructions For Use Manual

Anesthesia workstation
Hide thumbs Also See for Fabius plus:
Table of Contents

Advertisement

Troubleshooting

Ventilator failure

Alarm VENTILATOR FAIL !!!

If the ventilator does not return to its initial state,
the alarm VENTILATOR FAIL !!! activates.
Only manual ventilation or spontaneous breathing
is possible.
No other ventilation modes can be selected.
In this case, proceed as follows:
1 Change to ManSpont ventilation mode.
2 Set the APL valve to position Man.
3 Set the APL valve to the desired pressure.
4 Fill the breathing bag, if necessary with the aid
of the O
flush key.
2
5 Manually ventilate the patient.
Bypassing the ventilator
In the following cases, the ventilator must be
bypassed so that the ventilation can be continued.
– The ventilator does not return to its initial state
after a malfunction.
and
– The spontaneous breathing mode cannot be
activated.
To bypass the ventilator, proceed as follows:
1 Set the On/Off switch on the rear of Fabius to
(off).
2 Set the On/Off switch back to
158
Fabius restarts and performs a selftest. More
information on the selftest can be found in section
''Checking the readiness for operation'' on page
88.
1 Select ventilation mode ManSpont.
2 Set the APL valve to position Man.
3 Set the APL valve to the desired pressure.
4 Fill the breathing bag, if necessary with the aid
of the O
5 Manually ventilate the patient.
Before starting the ventilation with an automatic
ventilation mode, contact DrägerService or the
authorized local service partner.
(on).
flush key.
2
Instructions for use Fabius plus SW 3.n