Leakage Current On Applied Part With Test Adapter (Normal Condition) - Dräger Savina 300 Technical Documentation Manual

Intensive care ventilator
Hide thumbs Also See for Savina 300:
Table of Contents

Advertisement

NOTE
The test must be performed twice! The second test is performed with the
plug rotated 180° in the socket. In many test devices the mains plug rotation
is simulated by means of a built-in selector switch. The higher measured val-
ue must be documented.
NOTE
The reference value (initial measured value) must always be transmitted!
NOTE
If the measured values are between 90% and 100% of the permissible limit
value, the reference value and the previously measured values of the recur-
rent test should be applied to assess electrical safety!
Test
The reference value must not exceed 5000 μA.
Result
Reference value
Test
The recurrent test value must not exceed 5000 μA.
Result
Recurrent test
3.1.7

Leakage current on applied part with test adapter (normal condition)

NOTE
The following measurement is performed under „normal condition".
Prerequisites
The tester is switched on.
Test set-up
Fig. 4
Item Designation
1
2
3
4
5
L
N
PE
Leakage current on applied part
Tester
Device under test
Applied part of device under test
Test adapter
Configurable sockets for applied parts
Conductor
Neutral conductor
Protective earth conductor
Electrical safety
Electrical safety to EN 62353
[________μA]
[________μA]