Setting The Anesthetic Gas Concentration - Dräger Fabius plus Instructions For Use Manual

Anesthesia workstation
Hide thumbs Also See for Fabius plus:
Table of Contents

Advertisement

Operation

Setting the anesthetic gas concentration

C
D
E
Prerequisite: The vaporizer is mounted in
accordance with the instructions for use of the
respective vaporizer.
WARNING
Risk of patient injury
When the vaporizer is in the control dial
position T, has heated up due to high ambient
temperature, and is then used, a high
anesthetic gas concentration might be
delivered.
After connecting the vaporizer to the
anesthesia workstation, always turn the
control dial of the vaporizer to position 0 and
wait at least 15 seconds to enable a pressure
equalization.
The following section describes the operation of
Vapor 2000.
 Set the fresh-gas flow on the anesthesia
workstation.
92
If the control dial is in position T:
1
B
2
A
3 Check the filling level on the sight glass (C)
4 When the refill mark (D) is reached, 250 mL
5 Fill the vaporizer at the latest when the
Press the 0 key (A) and engage the control
dial (B) to position 0. To ensure pressure
equalization, wait at least 15 seconds.
Press the 0 key (A) and set the control dial (B)
counterclockwise to the desired anesthetic gas
concentration.
regularly. The filling level must be between
minimum and maximum.
(normal anesthetic agent bottle) can be refilled.
minimum mark (E) is reached, refer to
instructions for use of the respective vaporizer.
Instructions for use Fabius plus SW 3.n