Configuration
Results of the system leakage test
The test of the system leakage can have the
following results:
System leakage
[mL/min]
250
251 to 350
>350
Test of the overpressure safety valve
This test checks the functionality of the
overpressure safety valve.
The test results are displayed on the screen with
the leakage test results (B).
136
Displayed result
[mL/min]
Measured value and
PASSED
Measured value and
FAILED
>350 and FAILED
Troubleshooting with failed test of the
overpressure safety valve
WARNING
Risk of unexpected occurring overpressure
A soiled or non-functioning overpressure
safety valve is not able to compensate for
suddenly occurring overpressure in the
breathing system.
Perform the leakage test before start-up of
the device. Observe the test results of the
overpressure safety valve.
 Repeat leakage test. If the test of the
overpressure safety valve continues to fail,
contact DrägerService or the authorized local
service partner.
Instructions for use Fabius plus SW 3.n