Dräger Fabius plus Instructions For Use Manual page 199

Anesthesia workstation
Hide thumbs Also See for Fabius plus:
Table of Contents

Advertisement

Accuracy
Pressure limitation (PMAX)
Tidal volume (VT)
Respiratory rate (Freq)
Ratio of inspiratory time to expiratory time (TI:TE)
Inspiratory pause (TIP:TI)
Positive end-expiratory pressure (PEEP)
Inspiratory pressure (PINSP)
Overpressure safety valve
Underpressure safety valve (inlet valve for ambi-
ent air)
Minimum pressure limit
Measuring the system compliance
Instructions for use Fabius plus SW 3.n
±5 cmH
O (±5 hPa) of the setting
2
±5 % of the setting or 20 mL, depending on which
value is higher (discharged to atmosphere, no
compliance correction)
±1 bpm (±1 1/min) of the setting or ±5 %, depend-
ing on which value is higher
±5 % of the setting
±25 % of the setting
±2 cmH
O (±2 hPa) or ±20 % of the setting,
2
depending on which value is higher
±2 cmH
O (±2 hPa) or ±20 % of the setting,
2
depending on which value is higher
75 ±5 cmH
O (75 ±5 hPa)
2
-7.5 to -9 cmH
O (-7.5 to -9 hPa)
2
-9 cmH
O (-9 hPa)
2
0.2 to 6.0 mL/cmH
O (0.2 to 6.0 mL/hPa)
2
±0.2 mL/cmH
O (±0.2 mL/hPa) or ±10 % of the
2
actual compliance depending which value is
higher
Technical data
199