ManSpont
B
The following alarms can be activated or
deactivated on the ManSpont screen with the ON
/ OFF key (B):
– Apnoea Pressure (see chapter ''Pressure
alarms in automatic ventilation modes'' on
page 128)
– Volume Alarms (see chapter ''Volume alarms''
on page 127)
Instructions for use Fabius plus SW 3.n
Changing to manual ventilation
NOTE
In ManSpont mode, the apnea alarm time for
triggering apnea volume alarms is extended from
15 seconds to 30 seconds (Caution category)
and from 30 seconds to 60 seconds (Warning
category).
B
1 Set the APL valve head to the desired
maximum airway pressure.
Settings between the grid marks are also
possible.
A
2 Press the ManSpont (A) key.
3 Confirm the new mode.
10
20
Volume Control
Operation
97