Deactivating the O
Volume Control
If Fabius is configured by DrägerService for
operation with deactivated O
following functions of the O
deactivated:
– Parameter field for O
– Setting of O
alarm limits
2
– Calibration of O
sensor
2
– Alarm for the inspiratory O
the O
sensor
2
The message No Integrated O
displayed in the O
monitoring window (A).
2
NOTE
If the internal FiO
monitoring is deactivated, an
2
external FiO
monitoring must be available in
2
accordance with general safety standards.
Instructions for use Fabius plus SW 3.n
monitoring
2
A
monitoring, the
2
monitoring, are
2
monitoring
2
concentration and
2
Monitoring! is
2
Monitoring
125