Configuration

Restoring the default settings

The default settings are restored in the following
cases:
– Switching Fabius on and off
– Performing the system test
– Pressing the Restore Site Defaults softkey
Standby
1 Press the Restore Site Defaults softkey (A).
The default settings are restored. The following
message is displayed using the standby softkeys
(B):
Site Default settings restored
The default settings can be adjusted on the
Standby Set-up screen. Adjustment of the default
settings is password-protected.
WARNING
Risk due to unsuitable ventilation settings
After the default settings are restored, check
whether the settings for the ventilation and
monitoring are suitable for the patient.