Troubleshooting
Alarm
Alarm
priority
(23/31)
APNOEA PRESSURE !!!
(20)
APNOEA VENTILATION !!
(7)
BATTERY LOW !
(17)
BATTERY LOW !!
164
Cause
The apnea pressure alarm is
based on a time staggering.
In the Volume Control,
Pressure Control, SIMV/PS
modes with Freq ≥ 6 or in
Pressure Support mode
with apnea ventilation deacti-
vated:
Warning = PAW did not
exceed the pressure thresh-
old value for a duration of >30
seconds.
In the ManSpont, SIMV/PS
modes with Freq < 6 or in
Pressure Support mode
with apnea ventilation acti-
vated:
Warning = PAW did not
exceed the pressure thresh-
old value for a duration of >60
seconds
Breathing/ventilation
stopped.
Leakage or disconnection in
the breathing system.
Breathing/ventilation
stopped.
Leakage or disconnection in
the breathing system.
If two or more successive
breaths of the apnea ventila-
tion are automatically trig-
gered, the settings for
Pressure Support are not
correct.
No mains power and battery
<20 %
No mains power and battery
<10 %
Instructions for use Fabius plus SW 3.n
Remedy
Check ventilator.
Check breathing system.
Check ventilator.
Check breathing system.
Fabius detects a sponta-
neous breath of the patient.
Check settings for Pres-
sure Support.
Restore mains power sup-
ply.
Restore mains power sup-
ply.