Annex
Checklist
The checklist for daily checkout before using the
medical device takes into consideration all
possible configurations of Fabius. If a test point
does not apply to the Fabius to be tested due to
configuration deviations, skip the respective test
point.
All checks must be performed daily before use of
the device. Personnel who perform the checks
must be fully familiar with the instructions for use.
P
This test point must be performed before
each patient change.
Box to check if the test point passed.
Mark off the individual functions after successful
check.
Make copies of these pages so that they can be
used as a daily record for the device check.
Fabius plus
Part number:
Prerequisites
The maintenance intervals of the device
and accessories have not been exceeded.
P
The device is completely assembled and
connected.
All monitoring functions (e.g., O
ing) and external monitors (e.g., breathing
gas monitor) are switched on and function-
ing.
The system test for Fabius has been per-
formed.
P
The sample line for gas monitoring (if
present) is connected to the Luer Lock
connection on the Y-piece.
P
An appropriate anesthetic agent was
selected.
P
D-Vapor (if present) is switched on.
220
Battery
P
Make sure that the battery is completely
charged. (Battery operation of 45 minutes
is only guaranteed with a completely
charged battery.)
Gas supply
Visually inspect the central gas supply and
the gas cylinders. Connect all hoses.
Makes sure that the hoses are tightly con-
nected.
Make sure that the supply pressures of the
central gas supply lie within the permissi-
ble range.
Gas cylinders
Open the gas cylinders (if present).
O
(70 kPa x 100).
N
(43 kPa x 100).
Air pressure is higher than 1000 psi
(70 kPa x 100).
Close the gas cylinders.
monitor-
2
O
flush
2
Press the O
strong gas flow escapes from the patient
connection on the Y-piece.
Release the O
gas flow stops.
pressure is higher than 1000 psi
2
O pressure is higher than 600 psi
2
flush key: Check whether a
2
flush key: Check that the
2
Instructions for use Fabius plus SW 3.n