Testing/Calibrating the CO
The CO
sensor is factory calibrated and can be used
2
without further calibration on any Evita 2 dura ventilator.
A CO
calibration is performed as part of the ventilator
2
check.
Calibration of the CO
sensor is only required:
2
– if, upon checking calibration with a test filter or
with test gas, the specified test values are not met
– during the half-yearly preventive maintenance
inspections.
The calibration check or calibration may be performed
during ventilation.
Testing CO
Calibration With Test Filter
2
Perform CO
calibration, see pages 11,12.
2
After the CO
calibration:
2
Press »Calib./Config.« menu key.
Using the »Sensor jj« menu key, select »CO2 Test«
menu page.
Display (example):
Activate »Measure« screen key = press dial knob.
Sensor
2
Testing/Calibrating the CO
Testing CO
Calibration With Test Filter
2
Sensor
2