Overview

External fresh-gas outlet

The external fresh-gas outlet is used for
connecting different non-rebreathing systems, e.g.,
Magill, Kuhn, Waters, Bain.
WARNING
Risk of excessive airway pressure
Without a pressure-relief valve or breathing
bag, airway pressure may become too high.
Only connect non-rebreathing systems with
breathing bag or pressure-relief valves that
comply with applicable safety standards.
WARNING
Insufficient gas supply to the patient
Non-rebreathing systems are only intended
for manual ventilation or spontaneous
breathing and must only be connected to the

external fresh-gas outlet.

When using a non-rebreathing system,
ensure an adequate gas monitoring.
WARNING
Risk of misinterpretation of measured values
The values for O
, pressure, and volume
2
displayed on Fabius do not correspond to the
values for the patient connected to external
fresh-gas outlet as they are based on
measurements taken at the compact
breathing system.
When using the external fresh-gas outlet,
change into the Standby mode.
34
WARNING
Risk of faulty gas delivery
O
and CO
and any anesthetic gases must
2
2
also be monitored for non-rebreathing
systems.
The sample line must be connected with the
connector of the non-rebreathing system and
the connector of the gas analyzer (e.g., Scio,
Vamos).
NOTE
Strictly observe the instructions for use of the
non-rebreathing system (e.g., Magill, Waters,
Bain).
Instructions for use Fabius plus SW 3.n