For your safety and that of your patients
WARNING
Risk of crushing
If the writing tray is not correctly locked in
place, objects can fall down or fingers and
breathing hoses, for example, can be
pinched.
Make sure that the writing tray is correctly
locked when folding down or sliding into the
device.
WARNING
Risk of crushing
Movable device parts or attached
components may cause crushing due to
clamping. Pay special attention to edges,
movable parts, and corners when working
with the following components:
– Breathing system cover
– Drawers
– Extensible writing tray
– Swivel arms for mounted devices
– Accessories such as gas cylinders,
vaporizers, CLIC absorber, and CLIC
adapter
14
WARNING
Risk of electric shock
This device is only intended for use in rooms
in which the power lines correspond to the
national applicable safety standards for
patient's rooms in hospitals. Observe the
following points to avoid electric shock:
– The covers of the components must not
be removed.
– Maintenance work must only be
performed by DrägerService. Use only
grounded electrical connections and
power cables that meet hospital
standards.
– Before connecting the medical device,
make sure that external devices are
grounded to meet the hospital standard (in
accordance with national applicable
regulations).
– Before cleaning work or maintenance
work is performed, disconnect all plugs
for the power supply.
– If the medical device has come in contact
with liquids, let it dry completely before it
is reconnected to the power supply.
– Check that the power cable is securely
clamped to the power inlet.
– Only connect additional devices if they
have been approved by Dräger.
WARNING
Risk of device failures
If the anesthesia workstation is used in a
tipped position, parts can be damaged or
their function can be comprised.
Do not use the anesthesia workstation at an
inclination over 1°.
NOTE
The device software of Fabius must be installed
by experts. Dräger recommends having the
software installation performed by
DrägerService.
Instructions for use Fabius plus SW 3.n