Ending Operation - Dräger Fabius plus Instructions For Use Manual

Anesthesia workstation
Hide thumbs Also See for Fabius plus:
Table of Contents

Advertisement

Operation

Ending operation

1 Press the
button (A) and confirm.
Ventilation monitoring and alarm monitoring are
switched off. The ventilator stops.
2 Set the control dial (B) of the vaporizer until it
engages on 0.
114
A
B
WARNING
Risk of material damage and risk to health
A vaporizer must never remain in operation
without a fresh-gas flow. High concentrations
of the anesthetic gas flow into the ventilation
circuit and the ambient air. This can lead to
material damage and contamination of the
ambient air with anesthetic gases.
Never cut off the fresh-gas flow before the
vaporizer is switched off.
3 Close the flow control valves.
Power-saving mode is activated after 2.5
minutes.
4 Close the cylinder valves.
NOTE
Leave Fabius connected to the mains power
supply so that the battery is not discharged
Instructions for use Fabius plus SW 3.n