Series 200a Fi User's Guide
84.
Shortly after xenon lamp is turned off, it cannot be turned on again, since it is still at a high temperature.
NOTE:
Therefore, measurement is impossible.
For re-measurement, wait at least 5 minutes before restarting the lamp.

Setting the GLP Function

For this function, the following items are settable.
: The logbook data entry is reset when a numeric value representing "year" is input followed by pressing
NOTE
the ENT key.
Even when the same year as desired is already indicated, it must be reentered without fail.
1. Display of Xe lamp LOGBOOK data
Total ON period of the Xe lamp and its standard lifetime (in parentheses under Xe LAMP), the number
of switching-on times and the date of the last lamp change can be indicated.
a. Press the GLP key.
b. The following display will be presented.
<GLP initial screen>
c. The following display will be presented.
<Lamp selecting screen>
d. The following display will be presented.
<1st monitor screen>
TIME
0.0
The Xe lamp comes on when this unit is powered on.
Display and resetting of Xe lamp LOGBOOK data
Display and resetting of Hg lamp LOGBOOK data
Setting of KEY LOCK
WL CHECK (wavelength accuracy check) by using Hg lamp
CHECK (Xe lamp energy check)
GLP (LOGBOOK=1, KEY LOCK=2,
1
WL CHECK=3, Xe CHECK=4)
Select LOGBOOK.
Press numeric key 1 and then ENT key.
LAMP SELECT (Xe=1, Hg=2)
1
Select Xe.
Press numeric key 1 and then ENT key.
Ex
Em
FL
250
350
0.0
PROG
LAMP
1
OFF
[I]