Series 200a Fi User's Guide
74.
WARNING
1: Dark → 7: Bright
Default value: 4
5. LCD contrast setting is now completed and the setup initial screen appears.
Under CONTRAST, the selected contrast value is indicated.
<Setup initial screen>
When you press the ENT key here, the setup items selecting screen is presented to allow
reentry.
The selected setting will be fetched upon turning on power supply at the next time.
6. Turn OFF the power switch.
Setting of Standard Xenon Lamp Lifetime (XE EXCHANGE
STANDARD TIME)
The standard life time of the xenon lamp is settable.
This setting activates the following function. If the total operation time of the Xe lamp is longer than the
standard lifetime upon turning on the lamp, an error message is indicated to notify that the time point of
lamp replacement is reached.
1. Turn on the power switch with the SET UP [UTLTY] key held down.
2. The following display will be presented.
<Setup initial screen>
Press the ENT key.
3. The following display will be presented.
<Setup items selecting screen>
Injury due to Xenon Lamp Explosion
Replace the xenon lamp with a new one before its cumulative turn-on time
exceeds the guaranteed useful lifetime.
With the lapse of the guaranteed useful lifetime, the bulb wall darkens due
to evaporation of electrode material and adhesion of scattered
substances, resulting in a decrease in heat dissipation. This may increase
temperature (pressure) inside the bulb excessively to cause explosion..
SETUP:
DETECTOR
CH1
SETUP:
DETECTOR
CH1
CONTRAST
Serial No.
3
CONTRAST
Serial No.
4