Assembling; Removal Of Transport Screws - PerkinElmer 200A Series User Manual

Fluorescence detector
Table of Contents

Advertisement

Series 200a Fi User's Guide
60.

Assembling

Removal of Transport Screws

1. Remove three screws from the bottom face.
Use a Phillips screwdriver to remove transport screws, *A, *B and *C
Transport screw *A
(for main body)
Fig. 4-1 Transport Screws on the Bottom Face
2. Remove the lamp house from the instrument. (See Chapter 6, Maintenance.)
3. Use a Phillips screwdriver to remove the transport screw
from the lamp house.
4. Mount the xenon lamp and place it into the unit. (See Chapter 6, Maintenance.)
5. Adjust the position of the xenon lamp.
Please adjust VERTICAL first because the lamp position is shifted downward for fixation of the
transport screw. (See Chapter 6, Maintenance.)
Transport screw *C (for lamp house)
Transport screw *B (for
main body)
*D
*E
,
and Transport TY RAP
*F
(Fig. 4-2)