Edit the Wavelength Program

If you need to edit the wavelength program follow the next procedure.
Press PROG key.
SET PROGE NO. (1 ~ 9)
1
Press 1 key and then ENT.
EDIT MODE (MODIFY=0, NEW=1)
0
Press 1 key and then ENT.
TIME
Ex
Em
0.0
250
350
Press 0 key and then ENT.
Press 2 5 0 keys and then ENT.
Press 3 5 0 keys and then ENT.
Press 1 key and then ENT.
Press 3 key and then ENT.
TIME
Ex
Em
0.0
250
350
Press 1 0 keys and then ENT.
Press 2 7 0 keys and then ENT.
Press 3 7 0 keys and then ENT.
Press 2 key and then ENT.
BASE
PMT
1
2
BASE
PMT
1
3
Operation
105