Series 200a Fi User's Guide
72.
Setup
Before analytical operation, communication channel (CH) and liquid crystal display contrast (LCD
CONTRAST) need to be set.
•
This setting is not required every time. When unnecessary, skip this setting.
•
The default value or currently set value is indicated before your input.
Selecting the Communication Channel (CH)
Set a communication channel via the E-Line in the following way.
1. Turn on the power switch with the SET UP [UTLTY] key held down.
2. The following display will be presented.
<Setup initial screen>
Press the ENT key.
3. The following display will be presented.
<Setup items selecting screen>
Select DETECTOR.
Press numeric key 1 and then ENT key.
4. The following display will be presented.
<Channel selecting screen>
Select a channel (CH).
Press numeric key 1 or 2 and then ENT key.
Default value: 1
Two channels can be selected for two detectors.
5. Channel setting is now completed and the setup initial screen appears.
Under DETECTOR, the selected channel setting is indicated.
SETUP:
DETECTOR
CH1
SETUP ITEMS (DETECTOR=1, CONTRAST=2
1
EXCHANGE STANDARD=3)
DETECTOR (CH1=1, CH2=2)
1
CONTRAST
Serial No.
4