28.

Environmental Conditions

Normal Operating Conditions:
Indoors use.
The location must be free of smoke, dust, corrosive fumes, direct
sunlight and excessive vibration.
Ambient temperature: +10°C to +35°C (+50°F to +95°F).
Relative humidity: 25% to 80%, without condensation.
Altitude: in the range 0 m to 2,000 m (sea level to 6,562 feet).
Mains fluctuations not exceeding +/- 10% of the nominal voltage.
Safe Operating Conditions:
Indoor use.
Ambient temperature: of +5
Relative humidity: 25% to 80%, without condensation.
Altitude: 0 m to 2000 m.
Storage Conditions:
Ambient temperature: –20°C to +60°C (–4°F to +140°F).
Relative humidity: 25% to 80%, without condensation.
Altitude: in the range 0 m to 12,000 m (sea level to 39,370 feet).
If the equipment is used in a manner not specified herein, the
protection provided by the equipment may be impaired.
WARNING
o
o
C to +40
C (+41°F to +104°F).