Series 200a Fl User's Guide
132.
Press numeric keys 9 8 and then ENT.
The present offset value is indicated at Em WL ADJUST. Indicated at the bottom right of
the screen is the wavelength check result (deviation of wavelength; indication range of –10
to 10).
Wavelength calibration is conducted by entering a new offset value (value to which the
deviation is added).
Press numeric keys (from 1 to 20) and then ENT.
Input range for offset value: 1 to 20
Calculation of new offset value:
New offset value =
Example: 12 = 10 – (-2)

Check of Leak Sensor

1. Operational Check of Leak Sensor
a. Drip water onto the leak sensor.
Drip water
b. Confirm that a leak error is detected.
Specification: A leak error must be detected.
ii. The emission wavelength calibration mode screen appears.
present offset value – wavelength deviation
iii. The 1st monitor screen now appears.
Leak sensor