Error Messages - PerkinElmer 200A Series User Manual

Fluorescence detector
Table of Contents

Advertisement

160 Series 200a Fl User's Guide

Error Messages

Message
ROM ERROR
PRESS "CL" KEY TO CLEAR
MESSAGE.
RAM ERROR PRESS "CL" KEY
TO CLEAR MESSAGE.
PARAMETER ERROR
PARAMETERS ARE SET TO
DEFAULT VALUE.
LOG INFORMATION ERROR
LOG INFORMATION IS
INITIALIZED.
EPROM ERROR INSTRUMENT
INFO. IS INITIALIZED.
Ex (Em) SIDE OF WL DRIVE
MECHANISM ERROR
Ex (Em) SIDE ACCURACY
(254 nm) NOT DETECTED
LAMP OFF BY OVERHEAT!
PRESS "CL" KEY TO CLEAR
MESSAGE.
LAMP OFF BY EXTERNAL
INPUT PRESS "CL" KEY TO
TURN ON.
LAMP OFF BY EXTERNAL
ERROR PRESS "CL" KEY TO
TURN ON.
LAMP OFF BY LEAKAGE
PRESS "CL" KEY TO CLEAR
MESSAGE.
Description
ROM is faulty.
RAM is faulty.
Analytical parameters are
inadequate.
Log information is
inadequate.
Instrument parameters are
inadequate.
Failure in detecting Ex or Em
wavelength origin during
self-diagnosis test
(initialization).
Wavelength error exceeds
(10 nm during wavelength
accuracy check.
Abnormal temperature
detected.
Xe lamp turned off to
prevent its overheating.
Lamp has been turned off
due to an external contact
input. Busy status is
assumed.
Lamp has been turned off
due to an external error.
Busy status is assumed.
Liquid leakage has occurred.
Xe and Hg lamps have been
turned off. Busy status is
assumed.
Remedy
1) Press CL key, and 1st
monitor screen returns.
2) Contact our nearest
service center.
1) Press CL key, and 1st
monitor screen returns.
2) Contact our nearest
service center.
1) Press CL key, and 1st
monitor screen returns.
2) Turn on power again.
3) Contact our nearest
service center.
1) Press CL key, and 1st
monitor screen returns.
2) Turn on power again.
3) Contact our nearest
service center.
1) Press CL key, and 1st
monitor screen returns.
2) Turn on power again.
3) Contact our nearest
service center.
1) Cannot be released by key
operation.
2) Turn on power again.
3) Contact our nearest
service center.
Contact our nearest service
center.
1) Press CL key, and 1st
monitor screen returns.
2) Contact our nearest
service center.
1) Press CL key, and 1st
monitor screen returns.
2) Use UTLTY key to light up
the lamp again.
1) Release the key lock if it
has been activated.
2) Press CL key and light up
the lamp again.
1) Release the key lock if it
has been activated.
2) Press CL key and light up
the lamp again.
1) Press CL key, and 1st
monitor screen returns.
2) Use UTLTY key to light up
the lamp again.

Advertisement

Table of Contents

Troubleshooting

loading

Table of Contents