Operation Of Marker; Setting Of Time Program - PerkinElmer 200A Series User Manual

Fluorescence detector
Table of Contents

Advertisement

Series 200a Fi User's Guide
98.
<Error display>
e. The "ready for output" monitor screen will appear.
<"Ready for output" monitor screen>
Press the START key.
Pressing the ESC key prior to START key will cancel the process.
The output spectrum No. set before will be indicated at RCD NO.
f.
Pressing STOP key during spectrum output will cancel the process.
The output spectrum No. set before will be indicated at RCD NO.
g.
When the spectrum output is finished, the 1st monitor screen will return automatically.

Operation of Marker

A marker line is indicated only on the recorder. It is usable as an event marker to indicate, for example,
the start of measurement or the point where set conditions were changed during measurement.
Press the MARKER [1] key to indicate a marker line on the recorder data.

Setting of Time Program

A time program is used for automatic change of the chromatograph measurement conditions with time.
The time program takes priority over other settings.
The following items can be set.
Measurement time (TIME)
Excitation wavelength (Ex)
Emission wavelength (Em)
Baseline processing (BASE)
Photomultiplier voltage (PMT)
SPECTRA ARE MISMATCHED
PRESS "CL" KEY TO CLEAR MESSAGE
RCD NO.
SPEED
1
60
The output monitor screen appears, and a spectrum is output.
<Output monitor screen (spectrum output under way)>
RCD NO.
Ex
1
250
WL-RANGE (Em)
200-600 PRESS
Em
FL
SPEED
350
0.0
60
WIDTH:STD
"START"
MODE
Em

Advertisement

Table of Contents

Troubleshooting