During testing, if there is no SDS module, the alert box will appear with "No SDS
Module Detected, Continue? No/Yes'',after select "Yes", the testing will be
continued. As long as any one of the cut-off conditions is met, the test stops.
Stop running and re-running
During the running of battery test file, press [Shift]+[I-set](Function) composite
key, and the instrument interface displays the running state control interface of
the existing file. The user can press Stop to stop existing test and turn off the
List function. Or, press Reset to restart a new test.

5.3 Battery emulation function

When the IT-M3600 system is under the source mode, the user can emulate the
battery charge and discharge properties due to the unique current bipolarity de-
sign and variable output impedance. In the battery emulation function, the in-
strument can emulate the scenes where several batteries are in parallel
connection/serial connection, and set the initial state of charge of the battery
based on needs for true emulation of various states of the battery to meet vari-
ous test requirements of the user.
During the emulation of charge and discharge test functions of the battery under
the source mode, the user needs to set battery-related attributes, such as full-
state voltage value, empty-state voltage value, internal resistance value and ca-
pacity. Set the parallel connection and serial connection for emulation batteries,
as well as the positive and negative current limits of the battery group.
Batt
Battery emulation test function
Emula-
tor
Run
Recall
Run battery emulation function
Initial SOC
Set the initial state of charge of battery
Recall the battery emulation file to be run
Recall File=1/
Select the number of file to be recalled. The total number
10
of files is 10. When the selected file does not exist, the in-
strument would prompt that the file is empty.
Cell
Properties
Copyright © Itech Electronic Co., Ltd.
During Recall, the user can only view
the battery attribute, and cannot edit it.
Full Volt: Battery voltage value under full
state
Empty Volt: Battery voltage value under
empty state
Inner R: Battery internal resistance
value
Advanced Function