Usage method:
After the IT-M3600 instrument is connected to the anti-reverse connection mod-
ule, the instrument automatically detects the module connection state; if the
connection is succeeded, the SDS module setting in the system menu is de-
faulted as Enable, and the anti-reverse connection module is under enabled
state. Based on the user operating state, the module state requirements are as
follows:
After the anti-reverse connection module is connected, the user needs to con-
nect the remote measure terminal SENSE+/SENSE-; when the battery charge/
discharge test function (Function > Battery) is executed, the instrument auto-
matically sets Sense to On state in the system menu and detects the SENSE
state; if the connection fails or there is any abnormality, report the protection
prompt that SENSE fails to detect voltage. Press [Esc] to exit.
Prompt information for reverse connection protection. All prompt interfaces can
be exited by pressing the [Esc] key.
When the user conducts common test instead of battery charge/discharge
function, turn on the output, and the instrument panel prompts "SDS En-
abled". The user needs to set corresponding state of SDS in the system
menu and set the SDS module to Disable state. Otherwise, the instrument
output cannot be turned on.
When the user enters the battery charge/discharge mode (Function > Bat-
tery), the instrument automatically sets the anti-reverse connection module
SDS in the system menu to Enable state. Under this state, the anti-reverse
connection module can be disabled through manual or command mode.
No SDS Module detected: Under the battery charge/discharge mode, if the
prompt is given when the On/Off key on the front panel is pressed and SDS
is not inserted, the user can select to continue the execution or detection of
SDS module.
DC Port No Voltage: The prompt is given in the case of reverse connection
or abnormality at the DC terminal.
Sense Port No Voltage: The prompt is given in the case of reverse connec-
tion or abnormality at the Sense terminal.
Surge Suppressing: The prompt is given when the surge suppress is oper-
ated with command or when a battery with certain voltage is correctly con-
nected for surge suppress under the battery test mode.
Surge Suppress Done: The prompt is given when the surge suppress opera-
tion is succeeded.
Surge Suppress Fail: The prompt is given when the surge suppress opera-
tion is failed.