6.3.6 Setting the WARNING value for the pump N2 flow
SET POINT
N2 FLOW LOW?
Use the up and down arrow keys to change the setting value.
▲ ▼
▲ : Increase the setting value by 0.1 Pam
▼ : Decrease the setting value by 0.1 Pam
Upper limit 81.0 Pam
Lower limit 2.0 Pam
Factory setting 4.0 Pam
Reset value for WARNING set value +1.0 Pam
6.3.7 Setting the WARNING value for the back pressure (option)
SET ALARM SP
BACK PRES.?
Use the up and down arrow keys to change the setting value.
▲ ▼
: Increase the setting value by 0.5 kPa.
▲
: Decrease the setting value by 0.5 kPa.
▼
Upper limit : 20.0 kPa
Lower limit : 5.0 kPa
Factory setting : 10.0 kPa
WARNING reset value : Set value –2.0 kPa
PRG.
N2 FLOW LOW
ENTER
# # .# Pam
3
/s
3
/s
3
/s
PRG.
BACK PRES.SP
ENTER
##.# kPa
ENTER
3
/s
3
/s
3
/s
3
/s
ENTER
Use the keys, "▲" and "▼" to set.
EBARA
(43)
SETTING COMPLETED
Setting
done
CORPORATION