Radeye Lost Date And Time Settings; Radeye's Battery Status Is Lower Than Expected; History Data Shows The Wrong Time And Date - Thermo Scientific RadEye GX Operating Instructions Manual

Multi-purpose survey meter for external gm-detectors
Table of Contents

Advertisement

Open next the serial interface step window in the RadEye.EXE
software: Configuration / Com settings... menu. The used serial
port has to be selected next.
10.6

RadEye lost date and time settings

The date and time information is lost, each time the batteries are
removed or below the required voltage for operation. The sim-
plest way to program the date and time is to use the "synchro-
nize" function of the optional RadEye.EXE software. Please refer
to chapter 8.5.2. Else the current date and time can be set also
through the RadEye's menu "Settings" / "Set Date / Time"
10.7

RadEye's battery status is lower than expected

Please check if the correct battery type is selected under "set-
tings" / "Batt. type" and change the type in case. Rechargeable
batteries supply a voltage of 1.2V each, while Alkaline batteries
provide 1.5V.
Rechargeable batteries offer also a more steep drop from accepta-
ble battery voltage to "battery low" than Alkaline types.
10.8

History data shows the wrong time and date

See chapter 8.5.2.
RadEye GX, GX-L
Pi/Ff 14.09.2016
DB-074 E
10-5