Physical Inspection; Probe Maintenance; 10-5-6-1Probe Related Checks; 10-5-6-2Basic Probe Care - GE LOGIQ 100 PRO Technical Publication

Table of Contents

Advertisement

GE H
EALTHCARE
D
5305880-100, R
IRECTION
10-5-5

Physical Inspection

Step
Item
1
Labeling
Scratches &
2
Dents
3
Control Panel
Control Panel
4
Movement
Cables &
5
Connectors
Shielding &
6
Covers
7
External I/O
Table 10-11 Physical Checks
10-5-6

Probe Maintenance

10-5-6-1
Probe Related Checks
Step
Item
1
Probe Holder
2
Probes
Table 10-12 Probe Related Checks
10-5-6-2
Basic Probe Care
The system user manuals and various probe handling cards provide a complete description of
probe care, maintenance, cleaning and disinfection. Ensure that you are completely familiar with
the proper care of GE probes.
Ultrasound probes can be easily damaged by improper handling. See the User Manual and probe
care cards for more details. Failure to follow these precautions can result in serious injury and
equipment damage. Failure to properly handle or maintain a probe may also void its warranty.
Any evidence of wear indicates the probe cannot be used.
Do a visual check of the probe pins and system sockets before plugging in a probe.
TEE and Interoperative probes often have special considerations and individual probe user
manuals. For TEE and Interoperative probes also refer to their separate user manuals.
Page 10-10
Ap p r o v e d Do c u m e n t - 5 3 0 5 8 8 0 - 1 0 0 T PH_ r 3 . p d f Pa g e 1 8 1 o f 1 9 7
3
EVISION
Verify that all system labeling is present and in readable condition.
Inspect the console for dents, scratches or cracks.
Inspect keyboard and control panel. Note any damaged or missing items.
Verify ease of control panel (Operator I/O Panel) movement in all acceptable directions. Ensure that
it latches in position as required.
Check all internal cable harnesses and connectors for wear and secure connector seating. Pay
special attention to footswitch assembly and probe strain or bend reliefs.
Check to ensure that all EMI shielding, internal covers, air flow panels and screws are in place.
Missing covers and hardware could cause EMI/RFI problems while scanning.
Check all connectors for damage and verify that the labeling is good.
Clean probe holders. (they may need to be soaked to remove excess gel).
Thoroughly check the system probe connectors and remove dust from inside the connector sockets
if necessary. Visually check for bent, damaged or missing pins
Section 10-3 - Periodic Maintenance Schedule